๐ First external quality assurance of antibody diagnostic for SARS-new coronavirus
To confirm an infection with the new coronavirus (SARS-CoV) causing the severe acute respiratory syndrome (SARS) diagnostic assays for detection of SARS-CoV specific antibody are necessary. To evaluate the diagnostic performance of laboratories an external quality assurance (EQA) study was performed in 2004. Participating laboratories (9/20) correctly detected anti-SARS antibodies in serum samples without false positive results in an immunofluorescence assay. In contrast, only 4/13 laboratories detected most of the anti-SARS antibody positive samples without false positive results using enzyme immunoassays (EIA) and/or immunoblot. The overall results clearly demonstrate that serological diagnosis of SARS-CoV remains at an early stage of development, with further technical improvements required, particularly with respect to the use of SARS specific EIAs.
keywords
๐ severe acute (1373)
๐ serum samples (106)
๐ respiratory syndrome (2004)
๐ acute respiratory (1734)
author
๐ค Niedrig, Matthias
๐ค Leitmeyer, Katrin
๐ค Lim, Wilina
๐ค Peiris, Malik
๐ค Mackenzie, John S.
๐ค Zambon, Maria
year
โฐ 2005
issn
๐ 13866532
volume
34
number
1
page
22-25
citedbycount
4
download
๐ [BibTeX]